TESTING THE STRENGTH OF THE ADHESIVE CONNECTION IN SPECIMEN - MONOTONIC TENSILE AND SHEAR WITH UNDER VARIABLE LOAD


Abstract eng:
Ceramic implants can create original terms of occlusion in mouth. The results obtained for zirconium dioxide as the ceramic material used in implantology. Describes the technology of the blank (trabecular) and research glued connection of two trabecular zirconia 3M ESPE LAVA with batch number:508150. Performed with the help of two different dental adhesive properties glue named: 3M ESPE RelyX and Kerr Maxcem Elite. Monotonic tensile tests were carried out for two groups of 15 samples and the shear tests were adopted one group of 11 samples. The results obtained are characterized by high reproducibility in the sample geometry made of zirconia, as confirmed by the p Shapiro-Wilk test (p = 0.06). It showed no statistical differences in an attempt to monotonic tensile Student's t test (p = 0.608).

Contributors:
Publisher:
Brno University of Technology, Institute of Solid Mechanics, Mechatronics and Biomechanics, Brno
Conference Title:
Conference Title:
Engineering Mechanics 2017
Conference Venue:
Svratka, CZ
Conference Dates:
2017-05-15 / 2017-05-18
Rights:
Text je chráněný podle autorského zákona č. 121/2000 Sb.



Record appears in:



 Record created 2017-05-22, last modified 2017-05-22


Original version of the author's contribution in proceedings, page 1054, section BIO.: