Influence of position of middle ear prosthesis on mechanics of hearing


Abstract eng:
The analysis of the mechanics of the normal middle ear and of the middle ear with ossicular prosthesis is performed. The creation of the finite element model of the middle ear is described. The model of the normal ear includes external ear canal, tympanic membrane, ossicle chain with joints to oval window in the enter of cochley. The model of prosthesis simulates T-like type of the commercial product. The amplitude of the stapes footplate velocity and the amplitude of pressure excitations in cochley were evaluated for different manners of prosthesis fixation.

Contributors:
Publisher:
Institute of Theoretical and Applied Mechanics, AS CR, v.v.i., Prague
Conference Title:
Conference Title:
ENGINEERING MECHANICS 2006
Conference Venue:
Svratka (CZ)
Conference Dates:
2006-05-15 / 2006-05-18
Rights:
Text je chráněný podle autorského zákona č. 121/2000 Sb.



Record appears in:



 Record created 2014-11-12, last modified 2014-11-18


Original version of the author's contribution as presented on CD, , paper No. 156. :
Download fulltext
PDF

Rate this document:

Rate this document:
1